A comparison of techniques to optimize measurement of voltage changes in electrical impedance tomography by minimizing phase shift errors

Bayford, Richard ORCID logoORCID: https://orcid.org/0000-0001-8863-6385, Eadie, L., Fitzgerald, A. J. and Holder, David S. (2002) A comparison of techniques to optimize measurement of voltage changes in electrical impedance tomography by minimizing phase shift errors. IEEE Transactions on Medical Imaging, 21 (6) . pp. 668-675. ISSN 0278-0062 [Article] (doi:10.1109/TMI.2002.800577)

[img]
Preview
PDF
Download (239kB) | Preview

Abstract

In electrical impedance tomography, errors due to stray capacitance may be reduced by optimization of the reference phase of the demodulator. Two possible methods, maximization of the demodulator output and minimization of reciprocity error have been assessed, applied to each electrode combination individually, or to all combinations as a whole. Using an EIT system with a single impedance measuring circuit and multiplexer to address the 16 electrodes, the methods were tested on resistor-capacitor networks, saline-filled tanks and humans during variation of the saline concentration of a constant fluid volume in the stomach. Optimization of each channel individually gave less error, particularly on humans, and maximization of the output of the demodulator was more robust. This method is, therefore, recommended to optimize systems and reduce systematic errors with similar EIT systems.

Item Type: Article
Research Areas: A. > School of Science and Technology > Natural Sciences
A. > School of Science and Technology > Natural Sciences > Biophysics and Bioengineering group
ISI Impact: 8
Item ID: 499
Useful Links:
Depositing User: Repository team
Date Deposited: 21 Nov 2008 14:48
Last Modified: 26 Jun 2022 12:28
URI: https://eprints.mdx.ac.uk/id/eprint/499

Actions (login required)

View Item View Item

Statistics

Activity Overview
6 month trend
262Downloads
6 month trend
400Hits

Additional statistics are available via IRStats2.