Optimized breath detection algorithm in electrical impedance tomography

Khodadad, Davood, Nordebo, Sven, Müller, Beat, Waldmann, Andreas, Yerworth, Rebecca J., Becher, Tobias, Frerichs, Inez, Sophocleous, Louiza, van Kaam, Anton, Miedema, Martijn, Seifnaraghi, Nima and Bayford, Richard ORCID logoORCID: https://orcid.org/0000-0001-8863-6385 (2018) Optimized breath detection algorithm in electrical impedance tomography. Physiological Measurement, 39 (9) . ISSN 0967-3334 [Article] (doi:10.1088/1361-6579/aad7e6)

[img]
Preview
PDF - Final accepted version (with author's formatting)
Download (1MB) | Preview

Abstract

This paper defines a method for optimizing the breath delineation algorithms used in Electrical Impedance Tomography (EIT). In lung EIT the identification of the breath phases is central for generating tidal impedance variation images, subsequent data analysis and clinical evaluation. The optimisation of these algorithms is particularly important in neonatal care since the existing breath detectors developed for adults may give insufficient reliability in neonates due to their very irregular breathing pattern. Our approach is generic in the sense that it relies on the definition of a gold standard and the associated definition of detector sensitivity and specificity, an optimisation criterion and a set of detector parameters to be investigated. The gold standard has been defined by 11 clinicians with previous experience with EIT and the performance of our approach is described and validated using a neonatal EIT dataset acquired within the EU-funded CRADL project. Three different algorithms are proposed that are improving the breath detector performance by adding conditions on 1) maximum tidal breath rate obtained from zero-crossings of the EIT breathing signal, 2) minimum tidal impedance amplitude and 3) minimum tidal breath rate obtained from Time-Frequency (TF) analysis. As a baseline the zero crossing algorithm has been used with some default parameters based on the Swisstom EIT device. Based on the gold standard, the most crucial parameters of the proposed algorithms are optimised by using a simple exhaustive search and a weighted metric defined in connection with the Receiver Operating Characterics (ROC). This provides a practical way to achieve any desirable trade-off between the sensitivity and the specificity of the detectors. [Abstract copyright: © 2018 Institute of Physics and Engineering in Medicine.]

Item Type: Article
Additional Information: ** From PubMed via Jisc Publications Router.
Article number = 094001
Keywords (uncontrolled): Electrical Impedance Tomography, Receiver Operating Characteristics, breath detection, global optimisation, inspiration, lung imaging, respiratory system
Research Areas: A. > School of Science and Technology > Natural Sciences > Biophysics and Bioengineering group
Item ID: 24819
Notes on copyright: This Accepted Manuscript is © 2018 Institute of Physics and Engineering in Medicine.
This Accepted Manuscript is available for reuse under a CC BY-NC-ND 3.0 licence (https://creativecommons.org/licenses/by-nc-nd/3.0/) after the 12 month embargo period provided that all the terms of the licence are adhered to.
Useful Links:
Depositing User: Jisc Publications Router
Date Deposited: 21 Aug 2018 10:39
Last Modified: 29 Nov 2022 19:39
URI: https://eprints.mdx.ac.uk/id/eprint/24819

Actions (login required)

View Item View Item

Statistics

Activity Overview
6 month trend
386Downloads
6 month trend
301Hits

Additional statistics are available via IRStats2.