Influence of clearance on the time-dependent performance of the hip following hemiarthroplasty: a finite element study with biphasic acetabular cartilage properties

Li, Junyan ORCID logoORCID: https://orcid.org/0000-0003-4053-8334, Hua, Xijin, Jin, Zhongmin, Fisher, John and Wilcox, Ruth K. (2014) Influence of clearance on the time-dependent performance of the hip following hemiarthroplasty: a finite element study with biphasic acetabular cartilage properties. Medical Engineering and Physics, 36 (11) . pp. 1449-1454. ISSN 1350-4533 [Article] (doi:10.1016/j.medengphy.2014.05.014)

[img]
Preview
PDF - Published version (with publisher's formatting)
Available under License Creative Commons Attribution 4.0.

Download (2MB) | Preview

Abstract

Hip hemiarthroplasty is a common treatment for femoral neck fracture. However, the acetabular cartilage may degenerate after hemiarthroplasty leading to postoperative failure and the need for revision surgery. The clearance between the acetabular cartilage and head of the prosthesis is one of the potential reasons for this failure. In this study, the influence of joint clearance on the biomechanical function of a generic hip model in hemiarthroplasty was investigated using biphasic numerical simulation. Both a prolonged loading period of 4000 s and dynamic gait load of 10 cycles were considered. It was found that a larger clearance led to a higher stress level, a faster reduction in load supported by the fluid and a faster cartilage consolidation process. Additionally, the mechanical performance of the acetabular cartilage in the natural model was similar to that in the hemiarthroplasty model with no clearance but different from the hemiarthroplasty models with clearances of 0.5mm and larger. The results demonstrated that a larger clearance in hip hemiarthroplasty is more harmful to the acetabular cartilage and prosthesis heads with more available dimensions (i.e. smaller increments in diameter) could be manufactured for surgeons to achieve a lower clearance, and reduced contact stress in hemiarthroplasty surgeries.

Item Type: Article
Research Areas: A. > School of Science and Technology > Natural Sciences > Biophysics and Bioengineering group
Item ID: 21905
Useful Links:
Depositing User: Junyan Li
Date Deposited: 05 Jun 2017 15:42
Last Modified: 29 Nov 2022 23:17
URI: https://eprints.mdx.ac.uk/id/eprint/21905

Actions (login required)

View Item View Item

Statistics

Activity Overview
6 month trend
150Downloads
6 month trend
301Hits

Additional statistics are available via IRStats2.