McBride, Jeffrey D. and Gabriel, Francis Guy and Fordham, John L. A. and Kolind, Torsten and Barcenas-Morales, Gabriela and Isenberg, David A. and Swana, Marlene and Delves, Peter J. and Lund, Torben and Cree, Ian A. (2008) Screening autoantibody profiles in systemic rheumatic diseases with a diagnostic protein microarray using a filtration-assisted Nanodot Array Luminometric ImmunoAssay (NALIA). Clinical Chemistry, 54 (5). pp. 883-890. ISSN 0009-9147

Full text is not in this repository.

Official URL: http://www.clinchem.org/cgi/content/full/54/5/883

This item is available in the Library Catalogue

Abstract

Background: We developed a cost-efficient modular system for multiplex analysis of the multiple autoantibodies that characterize systemic rheumatoid diseases. Methods: The nanodot array luminometric immunoassay (NALIA) system consists of conventional 96-well membrane-bottomed plates in which antigens or antibodies are adsorbed onto the underside of the membrane. Current arrays use a 5 x 5 format (25 dots/well), which allows 10 analytes to be measured in duplicate: double-stranded DNA (dsDNA), centromere protein B (CENP-B), PCNA, Sm, Sm ribonucleoprotein (Sm-RNP), U1-snRNP, Scl70, SSA/Ro, SSB/La, Jo-1, and controls. The test fluid, control sera, and subsequent reagents are drawn through the membrane. The captured analytes are quantified by monitoring chemiluminescence with a charge-coupled device (CCD) and analyzed with commercial array software. Results: The assay can detect <20 x 103 IU/L of anti-dsDNA. The interwell CV was 10%–14%. There was an 83% concordance ({kappa} = 0.56) between the NALIA results obtained for anti-dsDNA assayed by β-testing in a routine immunology diagnostic laboratory and the results obtained with a conventional ELISA reagent set. The concordance values for Ro, La, Sm, and RNP were 98% ({kappa}, 0.92), 93% ({kappa}, 0.41), 97% ({kappa}, 0.62), and 97% ({kappa}, 0.73), respectively. Conclusion: The NALIA approach promises to provide a highly economical platform for a wide range of applications that require assays of multiple analytes. The degree of concordance of our results with a conventional reagent set was no less than that occurring between different commercial products. A sample of serum from a finger stick provides a volume sufficient to perform the array assay.

Item Type:Article
Research Areas:School of Science and Technology > Natural Sciences
ID Code:3975
Deposited On:04 Feb 2010 15:23
Last Modified:22 Oct 2013 06:29

Repository staff only: item control page

Full text downloads (NB count will be zero if no full text documents are attached to the record)

Downloads per month over the past year