The use of hydroxyapatite-coated CAD-CAM femoral components in adolescents and young adults with inflammatory polyarthropathy: ten-year results

McCullough, C. J., Remedios, D., Tytherleigh-Strong, G., Hua, Jia and Walker, P. S. (2006) The use of hydroxyapatite-coated CAD-CAM femoral components in adolescents and young adults with inflammatory polyarthropathy: ten-year results. The Journal of bone and joint surgery. British volume, 88-B (7). pp. 860-864. ISSN 0301-620X

Full text is not in this repository.

Abstract

Between June 1991 and January 1995, 42 hydroxyapatite-coated CAD-CAM femoral components were inserted in 25 patients with inflammatory polyarthropathy, 21 of whom had juvenile idiopathic arthritis. Their mean age was 21 years (11 to 35). All the patients were reviewed clinically and radiologically at one, three and five years. At the final review at a mean of 11.2 years (8 to 13) 37 hips in 23 patients were available for assessment. A total of four femoral components (9.5%) had failed, of which two were radiologically loose and two were revised. The four failed components were in patients aged 16 years or less at the time of surgery. Hydroxyapatite-coated customized femoral components give excellent medium- to long-term results in skeletally-mature young adults with inflammatory polyarthropathy. Patients aged less than 16 years at the time of surgery have a risk of 28.5% of failure of the femoral component at approximately ten years.

Item Type: Article
Research Areas: A. > School of Science and Technology > Natural Sciences > Biophysics and Bioengineering group
Item ID: 16595
Useful Links:
Depositing User: Jia Hua
Date Deposited: 29 May 2015 14:24
Last Modified: 22 May 2019 10:02